FDAClass IIJanuary 24, 2024

GammaPod - Treatment Planning System-a component of the recall

GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System. Ref: XMSGP030A09.07-0.01 Intended for use in the noninvasive…

Why was GammaPod - Treatment Planning System-a component of the recalled?

Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

00850002615029(11)230321(21)TPS0018 (Distributed in the US) 00850002615029(11)230629(21)TPS0013 (Distributed in the US) 00850002615005(11)21H24(21)TPS0016 (Distributed in the US) 00850002615029(11)210629(21 )TPS0014 (Italy)

Product
GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System. Ref: XMSGP030A09.07-0.01 Intended for use in the noninvasive…
Recalling firm
Xcision Medical Systems, LLC
Quantity
4 units
Distribution
Worldwide - US Nationwide distribution in the states of MD, PA, TX and the country of Italy.
Recall date
January 24, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls