FDAClass IIDecember 24, 2014

PKG, 5MM X 33CM INSERT, DISSECTING FORCEPS, P/N recall

PKG, 5MM X 33CM INSERT, DISSECTING FORCEPS, P/N 0250080760 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and…

Why was PKG, 5MM X 33CM INSERT, DISSECTING FORCEPS, P/N recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

746657 845059 940773 941253 941254 941737 1041990 1042777 1042778 1042779 1241174 1241794 1244656 1245030 1245597 1246158 1340472 1341544 1341545 1342769 1343384 1343715 1345093 1441173 444723

Product
PKG, 5MM X 33CM INSERT, DISSECTING FORCEPS, P/N 0250080760 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls