Giraffe and Panda Bag and Mask Resuscitation System The recall
Giraffe and Panda Bag and Mask Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask…
Why was Giraffe and Panda Bag and Mask Resuscitation System The recalled?
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model M1139612
- Product
- Giraffe and Panda Bag and Mask Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask…
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 276 units total, 109 of those nationwide.
- Distribution
- Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the countries of Australia, Belgium, Canada, China, France, New Zealand, Poland, Saudi Arabia, Switzerland and United Kingdom.
- Recall date
- February 20, 2013
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)