FACSLyric 3L8C Instrument US-IVD recall
FACSLyric 3L8C Instrument US-IVD, Catalog Number 662877, UDI 00382906628770 The K170974 clearance document describes the device as follows: The BD…
Why was FACSLyric 3L8C Instrument US-IVD recalled?
On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted Catalog Number: 662877 UDI-DI code: 00382906628770 Instrument Serial Numbers / Power Supply Serial Numbers: Z662877000064 V23180197 Z662877000053 V23090166 Z662877000052 V23090305 Z662877000054 V23090341 Z662877000051 V23090060 Z662877000056 V23090037 Z662877000058 V23090446 Z662877000055 V23090190 Z662877000061 V23180242 Z662877000057 V23090456 Z662877000060 V23180454 Z662877000065 V23180453 Z662877000062 V23300158 Z662877000063 V23300110 Z662877000059 V23180441 Z662877000068 V23300012
- Product
- FACSLyric 3L8C Instrument US-IVD, Catalog Number 662877, UDI 00382906628770 The K170974 clearance document describes the device as follows: The BD…
- Recalling firm
- Becton, Dickinson and Company, BD Bio Sciences
- Quantity
- 16 power supplies
- Distribution
- Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Maryland, Massachusetts, Michigan, Minnesota, Nebraska, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia & Wisconsin. The countries of China, Costa Rica, Singapore, Thailand.
- Recall date
- January 1, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)