REF 0562K recall
REF 0562K (Adult Elongated O2/CO2 Mask - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.
Why was REF 0562K recalled?
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot # 081920N38 (Distributed outside US)
- Product
- REF 0562K (Adult Elongated O2/CO2 Mask - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.
- Recalling firm
- Westmed, Inc.
- Quantity
- 16,000 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom
- Recall date
- January 6, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)