Why was InPen Diabetes Management App recalled?
Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
InPen Software App for Android, versions 5.6.1.1 and 5.6.2 with InPen System REF/UDI-DI: MMT-105ELBLNA/10862088000313, MMT-105ELGYNA/10862088000320, MMT-105ELPKNA/10862088000337, MMT-105NNBLNA/10862088000344, MMT-105NNGYNA/10862088000351, MMT-105NNPKNA/10862088000368.
- Product
- InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA…
- Recalling firm
- Companion Medical, Inc.
- Quantity
- 140
- Distribution
- US Nationwide distribution including in the states of CA, NY, MT, OH, IN, MN, PA, CT, TX, WA, TN, SD, MI, FL, OR, NC, KY, WI, MO, WV, NV, SC, AZ, AR, MA, LA, VT, GA.
- Recall date
- December 28, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)