Nextra CH Cannulated Hammertoe Mini Instrument recall
Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser…
Why was Nextra CH Cannulated Hammertoe Mini Instrument recalled?
During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Number: 1578401 UDI: 00817701026389
- Product
- Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser…
- Recalling firm
- Nextremity Solutions
- Quantity
- 23 units
- Distribution
- US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.
- Recall date
- March 16, 2022
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)