FDAClass IIDecember 24, 2014

PKG, 5MM INSERT, PARK FORCEPS, P/N 0250080733 recall

PKG, 5MM INSERT, PARK FORCEPS, P/N 0250080733 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue…

Why was PKG, 5MM INSERT, PARK FORCEPS, P/N 0250080733 recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

541778 645498 746494 940777 940819 942547 943493 1040240 1040242 1041188 1041617 1042440 1042802 1042810 1042824 1042825 1141217 1141787 1142297 1144497 1144989 1145355 1145793 1146356 1146666 1240947 1241795 1242136 1245816 1246637 1342739 1344134 1344201 1344205 1344875 1345230 1345275 1345417 1345518 1345735 1440481 1441633 1441835 1442174 246637

Product
PKG, 5MM INSERT, PARK FORCEPS, P/N 0250080733 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)