FDAClass IIJanuary 1, 2020

STA Unicalibrator recall

STA UNICALIBRATOR (ref. 00675)

Why was STA Unicalibrator recalled?

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF: 00675 Lot Numbers: 253180, 253964, 254721, 255108 UDIs: (01)13607450006756(11)171231(17)191231(10)253180(241)00675 (01)13607450006756(11)180630(17)200630(10)253964(241)00675 (01)13607450006756(11)181130(17)201130(10)254721(241)00675 (01)13607450006756(11)190228(17)210228(10)255108(241)00675

Product
STA UNICALIBRATOR (ref. 00675)
Recalling firm
Diagnostica Stago, Inc.
Quantity
5,498
Distribution
US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK
Recall date
January 1, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls