FDAClass IIJanuary 24, 2024

PosiFlush Prefilled Saline Syringe with General Pump recall

PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547

Why was PosiFlush Prefilled Saline Syringe with General Pump recalled?

An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304

Product
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Recalling firm
Becton Dickinson & Company
Quantity
5,085,120 units
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.
Recall date
January 24, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)