PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N recall
PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating…
Why was PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N recalled?
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
941070 941263 941599 942144 943866 944196 1041986 1044516 1045038 1045226 1141655 1143142 1243797 1246161 1343131 1343403 1441956
- Product
- PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating…
- Recalling firm
- Stryker Endoscopy
- Quantity
- 180,573 units total
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Recall date
- December 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)