ExacTrac Dynamic software recall
ExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST; 20910-01A, ETD Positioning and Moni. 1.0.1 INST; 20910-01B…
Why was ExacTrac Dynamic software recalled?
The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Software versions 1.0.0, 1.0.1, 1.0.2, 1.0.3, GTIN 04056481142315.
- Product
- ExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST; 20910-01A, ETD Positioning and Moni. 1.0.1 INST; 20910-01B…
- Recalling firm
- Brainlab AG
- Quantity
- 205 systems
- Distribution
- Distribution was made to AZ, AR, CA, CO, FL, GA, IN, KY, MD, MS, MO, NJ, NY, NC, OK, OR, PA, TX, VA, and WA. There was no government or military distribution. Foreign distribution was made to Australia, Austria, Bahrain, Belgium, China, Denmark, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, New Zealand, Poland, Singapore, Slovakia, South Korea, Spain, Thailand, Turkey, United Kingdom, and United Arab Emirates.
- Recall date
- March 16, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)