Lumex, Patriot Full Electric Homecare Bed, & Patriot LX recall: Contamination
Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part…
Why was Lumex, Patriot Full Electric Homecare Bed, & Patriot LX recalled?
GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
BED SERIAL NUMBER RANGE US0458 / US0468 Patriot Full Electric 120V 697142145108 through 697142147927 US0458 / US0468 Patriot Full Electric 220V 697144140132 through 697144140172 US6000 Patriot LX Full Electric 120V 69204141262 through 69204141737 SERVICE MOTOR KITS WITH SHIP DATES OF MAY 4, 2015 THROUGH JUNE 30, 2016 Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912
- Product
- Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part…
- Recalling firm
- GF Health Products, Inc
- Quantity
- 3,547 (3063 beds, 484 motors)
- Distribution
- Worldwide Distribution - USA ( nationwide) Distribution including Puerto Rico and to the countries of : Bahamas, Canada, Chile, El Salvador and Santo Domingo.
- Recall date
- December 21, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)