FDAClass IIDecember 24, 2014

PKG, 5MM INSERT, RETENTION GRASPER, 45CM, P/N recall

PKG, 5MM INSERT, RETENTION GRASPER, 45CM, P/N 0250080725 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating…

Why was PKG, 5MM INSERT, RETENTION GRASPER, 45CM, P/N recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

940095 940142 943677 945463 1040531 1041187 1045225 1045651 1143646 1340474 1341991 1342737 1343404 1346101 1346660 1442042 1442251 1442789

Product
PKG, 5MM INSERT, RETENTION GRASPER, 45CM, P/N 0250080725 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)