FDAClass IIMarch 16, 2022

Hemashield Gold Knitted Microvel Double Velour Vascular recall: Labeling Error

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected…

Why was Hemashield Gold Knitted Microvel Double Velour Vascular recalled?

Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number: 21B10 S/N: 1318567653 UDI: (01)00384401017608(11)210210(17)260131(21)1318567653GS(10)21B10

Product
HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected…
Recalling firm
Maquet Cardiovascular, LLC
Quantity
1 unit
Distribution
US Nationwide distribution in the state of MO.
Recall date
March 16, 2022
Classification
Class II
Status
Terminated
Agency
FDA (device)