FDAClass IIMarch 16, 2022

1688 Camera Control Unit recall: Software Issue

1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic…

Why was 1688 Camera Control Unit recalled?

A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers of 1688 CCU, UDI: 07613327420081, with Software Revisions: 3.0.6, 4.0.13, and 4.0.18

Product
1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic…
Recalling firm
Stryker Corporation
Quantity
11,955
Distribution
Worldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL, IN, FR, ES, HK, IT, GB, CO, PE, MY, TH, CA, JP, MX, AR, NP, SG, PH, ZA, CL, KR, CN
Recall date
March 16, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)