FDAClass IIDecember 10, 2025

InSure ONE Fecal Immunochemical Test. Labeled as recall

InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.

Why was InSure ONE Fecal Immunochemical Test. Labeled as recalled?

InSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1. Model Number: 90010; GTIN:10850093003313; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 2. Model Number: 90025; GTIN: 20850093003310; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 3. Model Number: 90030; GTIN: 00850093003279; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102.

Product
InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.
Recalling firm
Enterix, Inc.
Quantity
201,426 test kits
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Nevada, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, Washington, West Virginia and the country of Philippines.
Recall date
December 10, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls