Why was MINC+ Benchtop Incubator recalled?
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316
- Product
- MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
- Recalling firm
- Cook Medical Llc
- Quantity
- 14 US; 101 OUS
- Distribution
- Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
- Recall date
- January 24, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)