FDAClass IIDecember 14, 2016

RayStation 4.0 recall: Software Issue

RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment…

Why was RayStation 4.0 recalled?

Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Build numbers: 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 or 5.0.2.35

Product
RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment…
Recalling firm
RAYSEARCH LABORATORIES AB
Quantity
US: 66 units
Distribution
US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington
Recall date
December 14, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)