FDAClass IIMarch 7, 2018

Allura Xper systems with Certeray X-ray Generator recall

Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended…

Why was Allura Xper systems with Certeray X-ray Generator recalled?

There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Allura Xper systems - Model #'s 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722063, 722064, 722065, 722066, 722078, 722079, (added as of 6/25/2018) 1120

Product
Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended…
Recalling firm
Philips Electronics North America Corporation
Quantity
705
Distribution
Worldwide Distribution - US Nationwide; foreign distribution to Brazil, Germany, Iceland, Japan, Mexico, and (updated 6/25/18) Saudi Arabia.
Recall date
March 7, 2018
Classification
Class II
Status
Terminated