76 cm (30") Appx recall: Contamination

76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN)…

Why was 76 cm (30") Appx recalled?

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lots: 4784390, 4784404, 4865931, 4865006, 4590283, 4774210, 4865952, 4711264, 4605516, 4723103, 4882016, 4806569 ,

Product
76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN)…
Recalling firm
ICU Medical, Inc.
Quantity
4790
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Recall date
December 30, 2020
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls