FDAClass IIDecember 30, 2020

5 IN (13 cm) Appx recall: Contamination

5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber…

Why was 5 IN (13 cm) Appx recalled?

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lots: 4726847 4764618 4749803 4882578 4719415 4858408 4763254 4799746 4909239 4763263 4807851 4849307 4711357 4734298 4784307 4877040 4603746 4751052 4809068 4719780 4887803 4902877 4741245 4773197 4793534 4853586 4603189 4749895 4800302 4749463 4751098 4775532 4853625 4877133 4902884 4726180 4751919 4763325 4742008 4763328 4877406 4902270 4589996 4741273 4751930 4764078 4849398 4894536 4742017 4784331 4894544 4774127 4719576 4787217 4908406 4752847 4866868 4908415 4854840 4853711 4894570 4759148 4605923 4784875 4800246 4807647 4866700 4590960 4802679 4889077

Product
5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber…
Recalling firm
ICU Medical, Inc.
Quantity
55000
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Recall date
December 30, 2020
Classification
Class II
Status
Ongoing
Agency
FDA (device)