Why was Panorama Patient Monitoring recalled?

An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

version 11.5.1

Product
Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit…
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Quantity
63 products
Distribution
Nationwide Distribution.
Recall date
February 10, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)