FDAClass IIDecember 24, 2014

PKG, 5MM INSERT, HOOK SCISSORS, 45CM, P/N 0250080573 recall

PKG, 5MM INSERT, HOOK SCISSORS, 45CM, P/N 0250080573 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…

Why was PKG, 5MM INSERT, HOOK SCISSORS, 45CM, P/N 0250080573 recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

542142 840087 840778 845037 848121 940157 943731 1040938 1041996 1042072 1042485 1042720 1046194 1141403 1145070 1242916 1242932 1243556 1244078 1344775 1344878 1345291 1345424 1440908 1441384

Product
PKG, 5MM INSERT, HOOK SCISSORS, 45CM, P/N 0250080573 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls