FDAClass IIIJanuary 16, 2019

Blazer Open-Irrigated Ablation Batch Catheter Product recall

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF)…

Why was Blazer Open-Irrigated Ablation Batch Catheter Product recalled?

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Blazer Open Irrigated Ablation Catheter Material Number: M00496200, Batch: 20399570, Expiration Date: 3/13/2020 Batch: 21778776, Expiration Date: 2/21/2021 Batch: 22152470, Expiration Date: 5/21/2021 Batch: 22499259, Expiration Date: 8/13/2001 Material Number: M0049620N40 Batch::20413170, Expiration Date: 3/16/2020 Batch: 20809676, Expiration Date: 6/22/2020 Material Number: M0049620K20 Batch: 20403945, Expiration Date: 3/13/2020

Product
Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF)…
Recalling firm
Boston Scientific Corporation
Quantity
1,282 (all three types in Japan)
Distribution
Japan
Recall date
January 16, 2019