FDAClass IIDecember 24, 2014

PKG, 5MM INSULATED SHAFT 33CM, FOR BIPOLAR FORCEPS, P/N recall

PKG, 5MM INSULATED SHAFT 33CM, FOR BIPOLAR FORCEPS, P/N 0250080371. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and…

Why was PKG, 5MM INSULATED SHAFT 33CM, FOR BIPOLAR FORCEPS, P/N recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

614716 714937 715668 715791 814961 816593 817443 817444 913326 913523 913761 913764 913951 913952 914226 914227 915061 1013608 1014813 1015780 1016062 1016400 1016825 1016827 1113018 1113019 1113020 1113339 1114541 1114542 1114543 1114795 1114796 1115296 1115297 1115651 1116119 1116258 1117011 1117012 1213123 1213124 1213125 1213731 1213732 1213978 1213979 1215123 1215124 1215125 1216148 1216149 1216234 1313047 1313943 1313944 1313945 1314475 1314476 1314696 1316718 1316803 1316804 1316805 1410179 1410180 1410181 1410182 1410415 1410416

Product
PKG, 5MM INSULATED SHAFT 33CM, FOR BIPOLAR FORCEPS, P/N 0250080371. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated