FDAClass IIJanuary 23, 2013

25-Hydroxy Vitamin D EIA recall

25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the quantitative…

Why was 25-Hydroxy Vitamin D EIA recalled?

Immunodiagnostic Systems is recalling the 25-Hydroxy Vitamin D EIA, AC-57F1 because the QC report was found to have an erroneous calibrator value assigned. The value of 24.4 ng/mL for calibrator 4 was stated on the QC report, however the correct value should have read 28.4 ng/mL.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot # 17628

Product
25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the quantitative…
Recalling firm
IDS (Immunodiagnostic Systems Ltd.)
Quantity
675 units
Distribution
US Nationwide Distribution
Recall date
January 23, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)