PROXIMATE¿ PPH Hemorrhoidal Circular Stapler recall
PROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE¿ PPH Hemorrhoidal Circular Stapler and Accessories have…
Why was PROXIMATE¿ PPH Hemorrhoidal Circular Stapler recalled?
Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Codes: P93K95 P9420Y P94A4T P93L1M P9420Z P94D3N P93M3G P9441H P94F3Z P93T35 P9450L P94G04 P93T7F P9463J P94H5F P93W3M P94765 P94J4W P93W3N P9487Z P94J80 P93Y58 P94901 P94K2A P94117 P94A20 R9200Z
- Product
- PROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE¿ PPH Hemorrhoidal Circular Stapler and Accessories have…
- Recalling firm
- Ethicon Endo-Surgery Inc
- Quantity
- 15,264 units
- Distribution
- Worldwide Distribution: US (Nationwide) and to countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, ECUADOR, EGYPT, INDIA, INDONESIA, JAPAN, KOREA, KUWAIT, LEBANON MEXICO, NICARAGUA, PAKISTAN, SAUDI ARABIA, SINGAPORE, TAIWAN URUGUAY and VIETNAM.
- Recall date
- January 16, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)