FDAClass IIDecember 24, 2014

PKG, NEEDLE HOLDER, STRIAGHT, P/N 0250080340. recall

PKG, NEEDLE HOLDER, STRIAGHT, P/N 0250080340. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue…

Why was PKG, NEEDLE HOLDER, STRIAGHT, P/N 0250080340. recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

247810 341537 440293 441251 441512 742141 744846 745483 840074 840509 846996 847695 847971 940051 940785 941250 942774 943684 944126 945441 1040422 1042501 1042937 1043498 1045032 1045646 1046195 1140658 1141568 1141927 1142299 1144210 1144626 1145149 1145983 1146678 1240546 1242127 1242255 1242935 1243538 1243953 1245819 1246065 1246178 1341532 1341556 1342750 1343397 1344034 1344462 1345425 1346314 1440438 1440466 1443606

Product
PKG, NEEDLE HOLDER, STRIAGHT, P/N 0250080340. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls