Moog Curlin Infusion recall
Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps.…
Why was Moog Curlin Infusion recalled?
Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015. Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.
- Product
- Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps.…
- Recalling firm
- MOOG Medical Devices Group
- Quantity
- 11,472
- Distribution
- Worldwide Distribution - US (nationwide), Guam, and countries of: Australia, Canada, Croatia, Turkey, New Zealand, Saudi Arabia, Japan, Korea, and Italy.
- Recall date
- February 17, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)