DePuy Spine EXPEDIUM Offset Torque Wrench recall
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset…
Why was DePuy Spine EXPEDIUM Offset Torque Wrench recalled?
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots
- Product
- DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset…
- Recalling firm
- DePuy Spine, Inc.
- Quantity
- 140 units
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
- Recall date
- January 23, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)