FDAClass IIDecember 24, 2014

PKG, STRONG FORCEPS, FENESTRATED, P/N 0250080320. recall

PKG, STRONG FORCEPS, FENESTRATED, P/N 0250080320. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…

Why was PKG, STRONG FORCEPS, FENESTRATED, P/N 0250080320. recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

840362 842132 847517 1040110 1040420 1042259 1042934 1045030 1045283 1046905 1141972 1143003 1144487 1145075 1145230 1145347 1147038 1240535 1242461 1243537

Product
PKG, STRONG FORCEPS, FENESTRATED, P/N 0250080320. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)