StatStrip Glucose Ketone Hospital Meter System recall: Software Issue
StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing
Why was StatStrip Glucose Ketone Hospital Meter System recalled?
A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model No 63683. UDI-DI (01)38548063685. Software versions: v0.0.13.10 to v0.0.13.44.
- Product
- StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing
- Recalling firm
- Nova Biomedical Corporation
- Quantity
- 347 units
- Distribution
- Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
- Recall date
- January 1, 2025
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)