FDAClass IIDecember 24, 2014

PKG, COBRA FORCEPS, P/N 0250080317. Laparoscopic Manual recall

PKG, COBRA FORCEPS, P/N 0250080317. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…

Why was PKG, COBRA FORCEPS, P/N 0250080317. Laparoscopic Manual recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

44074 343875 441699 845871 846764 847429 941590 944711 945214 945550 945806 945982 1040221 1040396 1041479 1042258 1042928 1045022 1140259 1140925 1142139 1142550 1143144 1144986 1145061 1145802 1146365 1146652 1240410 1243557 1244178 1244517 1341997 1342753 1343400 1343700 1344282 1440402 1440708 1441639 1441816 1443394 647436 647454

Product
PKG, COBRA FORCEPS, P/N 0250080317. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls