FDAClass IIJanuary 2, 2019

RayStation 3.5 recall

RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A

Why was RayStation 3.5 recalled?

Isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software version number: 3.5.0.16 , 3.5.1.6 , 4.0.0.14 , 4.0.1.4 , 4.0.2.9 , 4.0.3.4 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 5.0.0.37 , 5.0.1.11 , 5.0.2.35 , 6.0.0.24 , 6.1.0.26, 6.1.1.2 , 6.2.0.7 , 7.0.0.19 , 8.0.0.61 or 8.0.1.10

Product
RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A
Recalling firm
RAYSEARCH LABORATORIES AB
Quantity
69
Distribution
AZ CA CO CT DE FL GA HI ID IL LA MA MD ME MI MO MS NC NJ NY OH OR PA RI TN TX VA VT WA WI
Recall date
January 2, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls