FDAClass IIJanuary 27, 2016

GE Healthcare recall

GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It…

Why was GE Healthcare recalled?

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

AMTU01048 AMTU01023 AMTU00163 AMTU01050 AMTU00172 AMTU00175 AMTU00176 AMTU01034 AMTU01047 AMTU01055 AMTU01001 AMTU01002 AMTU01021 AMTU01044 AMTU01042 AMTU01015 AMTU00178 AMTU00179 AMTU00162 AMTU01000 AMTU01012 AMTU01041 AMTU00167 AMTU01019 AMTU00177 AMTU01049 AMTU00171 AMTU00164 AMTU01066 AMTU01006 AMTU00168 AMTU01061 AMTU01060 AMTU01056 AMTU01057 AMTU01073 AMTU00154 AMTU00166 AMTU00174 AMTU01005 AMTU01035 AMTU00150 AMTU01024 AMTU01025 AMTU01026 AMTU01027 AMTU01037 AMTU00152 AMTU01040 AMTU01022 AMTU00157 AMTU00158 AMTU00159 AMTU00160 AMTU00170 AMTU01018 AMTU01009 AMTU01032 AMTU01046 AMTU01045 AMTU01020 AMTU01008 AMTU00165 AMTU01013 AMTU01014 AMTU01033 AMTU01016 AMTU01036 AMTU01038 AMTU00151 AMTU00169 AMTU01003 AMTU00161 AMTU01010 AMTU01064 AMTU01011 AMTU01004 AMTU01017 AMTU01031 AMTU01053 AMTU01058 AMTU01059 AMTU01039 AMTU01007

Product
GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It…
Recalling firm
GE Medical Systems, LLC
Quantity
84
Distribution
Worldwide Distribution - US (nationwide including PR except NH, VT, & WY. Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.
Recall date
January 27, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)