FDAClass IIDecember 24, 2014

PKG, FUNDUS FORCEPS, P/N 0250080315. Laparoscopic recall

PKG, FUNDUS FORCEPS, P/N 0250080315. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…

Why was PKG, FUNDUS FORCEPS, P/N 0250080315. Laparoscopic recalled?

Stryker Endoscopy is recalling Laparoscopic manual instruments and accessories because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

141309 646670 842621 842622 842623 842624 848151 848152 848153 848154 848158 848159 944681 945330 945750 945828 945829 945830 945835 945836 1040394 1040541 1042480 1042926 1045981 1045982 1140693 1143447 1143582 1145552 1145982 1146297 1240548 1241307 1242395 1242465 1243545 1245973 1340289 1340717 1341251 1342684 1344461 134461 1345087 1345240 1345489

Product
PKG, FUNDUS FORCEPS, P/N 0250080315. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls