FDAClass IIDecember 7, 2016

MSS Cassette Glu/lac/urea recall: Software Issue

MSS CASSETTE GLU/LAC/UREA (BUN) as a part of the Roche OMNI S Analyzer: cobas b 221<6>Roche OMNI S6 system The Roche Diagnostics Omni S Analyzer is a…

Why was MSS Cassette Glu/lac/urea recalled?

Incorrect Urea results may be generated sporadically on cobas b 221<6>Roche OMNI S6 system and cannot be detected by QC. QC measurement results can be affected as well. This issue is only detectable if comparison result of a second measurement is available. The cause for this issue is an error in the software measurement procedure. In children, decreased urea results can point at inborn disorders or urea cycle. Therefore, false normal urea values could lead to delayed/incorrect diagnosis of such a disorder. Besides the patients most at risk, the risk for the overall patient population with falsely low or high values is considered remote. No severe adverse health effects are expected; but, unnecessary diagnostic measures may be initiated. Falsely normal results may result in further necessary examination not being carried out, with the possibility of a delay in the diagnosis and treatment. A medical risk cannot entirely be excluded.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number(s): All lots within shelf life. Expiration Date(s): All

Product
MSS CASSETTE GLU/LAC/UREA (BUN) as a part of the Roche OMNI S Analyzer: cobas b 221<6>Roche OMNI S6 system The Roche Diagnostics Omni S Analyzer is a…
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
97
Distribution
Nationwide Distribution including ALCA CO CT FL IL IN KS MD MI MO NC NE NJ NM NY OH OK PA PR SC TX VA and WI
Recall date
December 7, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)