FDAClass IIDecember 26, 2018

Reliance LITe Nerve Hook Bayoneted recall

Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction.…

Why was Reliance LITe Nerve Hook Bayoneted recalled?

Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot #s 183432, 183433, 183434, 17E168

Product
Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction.…
Recalling firm
Howmedica Osteonics Corp.
Quantity
39
Distribution
Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain.
Recall date
December 26, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls