(l) Visum Halogen Surgical Light recall
(l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the…
Why was (l) Visum Halogen Surgical Light recalled?
Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Code Number and Product Name: 0682000112-SPRING ARM, LIGHT 12-18KG, 0682000211 - FLAT PANEL SPRING ARM, ONDAL SUSPENSION, 15-21KG, 0682000212 - FLAT PANEL SPRING ARM, ONDAL SUSP., 12-18KG, 0682000226 - SPRING ARM,3.5-7KG,NAVIGATION, 0682000228 - SPRING ARM, LIGHT 9-15KG, 0682001062 - PKG EDS LIGHT SUSPENSION CENTRAL AXIS 5P/FP, 0682001270 - 9-POLE SPRING ARM, AC 2000,15-21 KG, 0682001274 - 5-POLE SPRING ARM AC 2000 15-21 KG, 0682400034 - FLAT PANEL SPRING ARM, EDS SUSPENSION, 15-21KG, 0682400059 - UNIVERSAL YOKE SPRING ARM RETROFIT KIT, 06824000591 - UNIVERSAL YOKE & SPRING ARM RETROFIT KIT, INT, 0682400151 - LOW CEILING SPRING ARM, 5P SIDE MOUNT, 0682400152 - LOW CEILING SPRING ARM, 9P SIDE MOUNT, 0682400169 - 5-POLE SPRING ARM AC 2000 9-15KG, CARDINAL, 0682400207 - STANDARD SPRING ARM WITH BEARING, 5P, 0682400208 - STANDARD SPRING ARM WITH BEARING, 9P, and 0682400365 - FLAT PANEL SPRING ARM WITH ERA ADAPTER BEARING.
- Product
- (l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the…
- Recalling firm
- Stryker Communications, Inc.
- Quantity
- United States: 1,082 units and Outside United States: 199 units
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Japan, Taiwan, Canada, United Kingdom, France, Netherlands, Spain, Mexico, India, Korea, Singapore, New Zealand, Hong Kong, Philippines and Brazil.
- Recall date
- January 16, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)