FDAClass IIJanuary 20, 2016

30 deg recall

30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high…

Why was 30 deg recalled?

This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model numbers: 470027-34, 470027-40, 470027-41. All lots.

Product
30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high…
Recalling firm
Intuitive Surgical, Inc.
Quantity
2624 total all models
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela. .
Recall date
January 20, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)