PKG, MICRO SCISSORS, STRAIGHT, P/N 0250080269. recall
PKG, MICRO SCISSORS, STRAIGHT, P/N 0250080269. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue…
Why was PKG, MICRO SCISSORS, STRAIGHT, P/N 0250080269. recalled?
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
842593 843071 843908 844480 1042087 1042653 1042654 1046703 1141794 1142621 1142918 1144316 1144618 1145408 1145972 1240118 1240977 1242391 1243271 1246632 1341583 1343296 1343401 1345285 1346226 1441560 1442811 153299
- Product
- PKG, MICRO SCISSORS, STRAIGHT, P/N 0250080269. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue…
- Recalling firm
- Stryker Endoscopy
- Quantity
- 180,573 units
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Recall date
- December 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)