FDAClass IIDecember 24, 2014

PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. recall

PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…

Why was PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

140280 143823 244698 341493 443333 541776 545061 647416 744587 842092 844289 846109 846772 846819 941064 941065 941852 942140 943501 943549 944300 945744 945886 945977 1040098 1040413 1041057 1041621 1041992 1042086 1042240 1042475 1042685 1042695 1042698 1042714 1044482 1045035 1045373 1045641 1046199 1046595 1046933 1046934 1046943 1142155 1142668 1142897 1142910 1143132 1144208 1144398 1144492 1144940 1145080 1145798 1145971 1146839 1146933 1240119 1241514 1241515 1242921 1244191 1244534 1245445 1246192E 1246638 1246638E 1246650 1340099 1341584 1341585 1343125 1343402 1343402E 1343705 1344204 1344286 1344456 1344880 1345236 1345532 1345532E 1345605 1345605E 1346432 1346666 1440405 1440684 1442518 1443219 1443653 1443917

Product
PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls