FDAClass IIFebruary 28, 2018

Pentax Video Duodenoscope recall

Pentax Video Duodenoscope Model: ED-3490TK (UDI of design being recalled: 04961333232420) These instruments are intended to provide optical…

Why was Pentax Video Duodenoscope recalled?

The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in order to mitigate the potential risk of infection in flexible endoscopy.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Serial Numbers

Product
Pentax Video Duodenoscope Model: ED-3490TK (UDI of design being recalled: 04961333232420) These instruments are intended to provide optical…
Recalling firm
Pentax of America Inc
Quantity
559
Distribution
USA (nationwide) Distribution to the states of including Puerto Rico : AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and WY, and the District of Columbia.
Recall date
February 28, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls