Vitros 5 recall: Software Issue
VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132, and VITROS 5,1 FS Chemistry System…
Why was Vitros 5 recalled?
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Software Version 2.8 and below; VITROS 5,1 FS Chemistry System Serial Numbers J34000016 - J34002323; VITROS 5,1 FS Chemistry System Refurbished Serial Numbers 34000133, 34000193, 34000197, 34000212, 34000285, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451, 34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557, 34000590, 34000596, 34000601, 34000631, 34000642, 34000761, 34000799, 34000805, 34000833, 34000851, 34000893, 34000946, 34000947, 34001006, 34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274, 34001297, 34001311, 34001317, 34001345, 34001378, 34001379, 34001389, 34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572, 34001612, 34001626, 34001633, 34001668, 34001755, 34001809, 34001840, 34001850, 34001875, 34001897, 34000123, 34000143, 34000191, 34000215, 34000219, 34000235, 34000251, 34000256, 34000260, 34000261, 34000319, 34000331, 34000380, 34000382, 34000406, 34000408, 34000412, 34000426, 34000431, 34000454, 34000464, 34000473, 34000501, 34000523, 34000530, 34000541, 34000542, 34000546, 34000548, 34000553, 34000558, 34000576, 34000577, 34000584, 34000585, 34000609, 34000612, 34000615, 34000616, 34000617, 34000618, 34000620, 34000625, 34000637, 34000644, 34000672, 34000744, 34000745, 34000755, 34000759, 34000776, 34000793, 34000800, 34000814, 34000831, 34000844, 34000846, 34000852, 34000856, 34000870, 34000876, 34000887, 34000892, 34000902, 34000912, 34000920, 34000938, 34000956, 34000969, 34000976, 34001007, 34001009, 34001044, 34001045, 34001052, 34001071, 34001075, 34001098, 34001099, 34001111, 34001129, 34001154, 34001161, 34001171, 34001182, 34001192, 34001195, 34001201, 34001204, 34001208, 34001212, 34001234, 34001247, 34001254, 34001271, 34001278, 34001284, 34001288, 34001289, 34001293, 34001298, 34001301, 34001302, 34001316, 34001325, 34001337, 34001340, 34001348, 34001367, 34001369, 34001399, 34001400, 34001407, 34001410, 34001416, 34001423, 34001427, 34001429, 34001457, 34001490, 34001518, 34001526, 34001551, 34001573, 34001580, 34001581, 34001582, 34001587, 34001589, 34001594, 34001610, 34001618, 34001623, 34001627, 34001673, 34001688, 34001693, 34001725, 34001738, 34001745, 34001747, 34001790, 34001796, 34001798, 34001810, 34001822, 34001831, 34001833, 34001849, 34001855, 34001859, 34001860, 34001867, 34001871, 34001872, 34001881, 34001905, 34001916, 34001923, 34001995.
- Product
- VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132, and VITROS 5,1 FS Chemistry System…
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- VITROS 5,1: Domestic - 932, Foreign - 1245; 5,1 Refurbished: Domestic - 63; Foreign - 153
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.
- Recall date
- January 20, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)