FDAClass IIFebruary 23, 2022

OER-Elite Endoscope Reprocessor-for use in cleaning and recall

OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and…

Why was OER-Elite Endoscope Reprocessor-for use in cleaning and recalled?

The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers. UDI: 04953170404047

Product
OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and…
Recalling firm
Olympus Corporation of the Americas
Quantity
295 units
Distribution
Nationwide
Recall date
February 23, 2022
Classification
Class II
Status
Terminated
Agency
FDA (device)