FDAClass IIDecember 23, 2020

Kowa VK-2s software Version 10.0.4.0 recall: Software Issue

Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated for…

Why was Kowa VK-2s software Version 10.0.4.0 recalled?

It has been determined that the software distributed with the retinal camera may delete patient information. Specifically, if multiple IDs are delated at the same time, their is a potential for all patient information to be deleted. This could result in delay of treatment.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 32168900007, 32168900008, 32168900009, 32168900010, 32168900011, 32168900019, 32168900020, 32168900021, 32168900022, 32168900025, 32168900035, 32168900036, 32168900037, 32168900044, 32168900045, 32169000052, 32169000144, 32169000145, 32169000152, 32169000153, 32169000166, 32169000167, 32169000168

Product
Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated for…
Recalling firm
Kowa Optimed Inc
Quantity
N/A
Distribution
US Nationwide distribution including in the states of NY, WA, TX, CT, MA, ID, IL, MI.
Recall date
December 23, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls