FDAClass IIDecember 24, 2014

PKG, OP-VIOPSY FORCEPS, SPOON JAW, P/N 0250080246. recall

PKG, OP-VIOPSY FORCEPS, SPOON JAW, P/N 0250080246. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…

Why was PKG, OP-VIOPSY FORCEPS, SPOON JAW, P/N 0250080246. recalled?

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

140616 245976 340614 344497 444951 540952 643595 846782 847704 940162 941321 942551 943424 943445 943536 943537 945451 1040934 1041474 1041845 1042234 1042681 1042691 1042707 1141159 1141977 1143134 1144330 1145150 1145799 1146352 1240419 1240501 1241813 1242394 1242476 1243369 1244536 1245518 1341465 1341598 1341600 1343695 1344290 1345041 1345100 1345606 1346109 1346670 1440677 1440707 1440894 1440896

Product
PKG, OP-VIOPSY FORCEPS, SPOON JAW, P/N 0250080246. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency