PKG, CLAW FORCEPS, P/N 0250080244. Laparoscopic Manual recall
PKG, CLAW FORCEPS, P/N 0250080244. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…
Why was PKG, CLAW FORCEPS, P/N 0250080244. Laparoscopic Manual recalled?
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
542920 744855 745487 747207 842094 846779 847453 940070 940790 942064 942847 943550 944301 945449 1040412 1042233 1042680 1042689 1042701 1045852 1046704 1141030 1141031 1142001 1142002 1144327 1144623 1145082 1145339 1145978 1146685 1146993 1147101 1240120 1240538 1240539 1241480 1241481 1242392 1242393 1242923 1242924 1244171 1244172 1245446 1245447 1246063 1246064 1341588 1341589 1341590 1343694 1346228 1346669 1440891
- Product
- PKG, CLAW FORCEPS, P/N 0250080244. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs…
- Recalling firm
- Stryker Endoscopy
- Quantity
- 180,573 units total
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Recall date
- December 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)