FDAClass IJanuary 27, 2016

arkray SPOTCHEM II Basic Panel - 1 Reagent test strips recall

arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin…

Why was arkray SPOTCHEM II Basic Panel - 1 Reagent test strips recalled?

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number PN5C26

Product
arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin…
Recalling firm
Arkray Factory USA, Inc.
Quantity
11 boxes (25 foil packaged test strips per box)
Distribution
Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
Recall date
January 27, 2016
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls